Comparison of adenoma detection rate using the novel 5-LED vs xenon-light endoscopic system: Propensity score matching analysis

 SFX Search Permissions and Reprints(opens in new window) Article preview thumbnailAbstract Background and study aims

Olympus’s new endoscopic system, EVIS X1, features five-LED illumination and a novel complementary metal-oxide-semiconductor (CMOS) image sensor distinct from conventional charge-coupled devices (CCDs), potentially improving colorectal adenoma detection rates (ADRs). This study compared ADR and related indicators between the EVIS X1 system and the conventional EVIS LUCERA ELITE, a xenon-light system.

Patients and methods

Of 4,915 colonoscopies performed between September 2020 and April 2023, 814 EVIS X1 and 953 LUCERA cases met inclusion criteria. After propensity score matching to balance baseline characteristics, 660 patients per group were analyzed. Outcomes included ADR, polyp detection rate (PDR), adenomas per colonoscopy (APC), and polyps per colonoscopy (PPC). Subgroup analysis assessed the impact of CMOS-equipped scopes within the X1 group.

Results

ADR was slightly higher in the X1 group (36.1%) than the LUCERA group (32.1%), although not statistically significant (P = 0.147). APC (0.77 vs. 0.61, P = 0.034) and PPC (0.95 vs. 0.75, P = 0.023) were significantly higher with X1. Within the X1 group, scopes with CMOS sensors achieved a significantly higher ADR (41.9%) compared with those without. Mean size of polyps detected was smaller with CMOS than with CCD scopes. Multivariate analysis identified age > 60 years, male sex, positive fecal occult blood test, and use of the X1 system with CMOS scopes as independent predictors of higher ADR.

Conclusions

The EVIS X1 system may have the potential to improve adenoma detection, particularly when used with CMOS sensor-equipped scopes. These findings suggest potential benefits for colorectal cancer screening, although further large-scale studies are warranted for validation.

Keywords Endoscopy Lower GI Tract - Polyps / adenomas / ... - Diagnosis and imaging (inc chromoendoscopy, NBI, iSCAN, FICE, CLE...) - Endoscopic resection (polypectomy, ESD, EMR, ...) Publication History

Received: 07 May 2025

Accepted after revision: 12 November 2025

Accepted Manuscript online:
01 December 2025

Article published online:
19 December 2025

© 2026. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).

Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany

Bibliographical Record
Tatsuhiro Ito, Satoshi Osawa, Takanori Yamada, Keisuke Inagaki, Tomohiro Takebe, Satoru Takahashi, Shunya Onoue, Kiichi Sugiura, Natsuki Ishida, Tomoharu Matsuura, Mihoko Yamade, Moriya Iwaizumi, Yasushi Hamaya, Ken Sugimoto. Comparison of adenoma detection rate using the novel 5-LED vs xenon-light endoscopic system: Propensity score matching analysis. Endosc Int Open 2025; 13: a27606529.
DOI: 10.1055/a-2760-6529