Simultaneous Estimation of Stavudine and Zidovudine in Tablet Dosage Forms by a Validated RP-HPLC Method for Quality Control Applications

Ankit Anchliya*, Rajesh Nagar, Sudha Vengurlekar and Sachin K. Jain

Faculty of Pharmacy, Oriental University, Sanwer Road Indore, M.P. India.

Corresponding Author E-mail:profankit.anchliya@gmail.com

Article Publishing History
Article Received on : 11 Jan 2025
Article Accepted on :
Article Published : 16 May 2025

ABSTRACT:

Using a UV-visible detector, a straightforward, repeatable, practical, and creative reversed-phase high performance fluid chromatographic technique was created and verified for the simultaneous measurement of stavudine and zidovudine in pharmaceutical measure forms. Acetonitrile: Water (50:50) was used as the mobile phase in the procedure, which was run on a C18 column at a run speed of 0.5 ml/min. 283 nm was the ideal wavelength for detection, yielding a superior drug detector response. 15 minutes was the runtime. The method's linearity, accuracy, precision, specificity, toughness, stability, boundary of quantification (LOQ), and limit of detection (LOD) were all verified. The suggested approach was effectively used for the instantaneous measurement of zidovudine and stavudine in pharmaceutical dose forms.

KEYWORDS:

Method Development; Method Validation; RP-HPLC; Stavudine; Zidovudine

Download this article as:  Copy the following to cite this article:

Anchliya A, Nagar R, Vengurlekar S, Jain S. K. Simultaneous Estimation of Stavudine and Zidovudine in Tablet Dosage Forms by a Validated RP-HPLC Method for Quality Control Applications. Orient J Chem 2025;41(3).


Copy the following to cite this URL:

Anchliya A, Nagar R, Vengurlekar S, Jain S. K. Simultaneous Estimation of Stavudine and Zidovudine in Tablet Dosage Forms by a Validated RP-HPLC Method for Quality Control Applications. Orient J Chem 2025;41(3). Available from: https://bit.ly/3HaFf2J


Introduction 

Antiretroviral treatment (ART) aims to achieve clinically meaningful immune reconstitution, which includes increases in naïve CD4+ cells and the reappearance of HIV- and other pathogen-specific lympho proliferative responses. The goal of the current investigation was to evaluate how zidovudine and stavudine sequencing affected the responses of CD4 lymphocytes and plasma HIV-1 RNA in individuals in the CHORUS observational cohort. Numerous individual analytical techniques are available for estimating the medications listed above in their separate dose forms or in combination with other dosage forms, according to the literature study 1-2. One dideoxynucleoside that is used to treat HIV infection is stavudine. The nucleoside overturn transcriptase inhibitor (NRTI) stavudine has capabilities. Phosphorylation of stavudine results in active metabolites that vie for viral DNA inclusion. They function as a sequence terminator of DNA combination and competitively block the HIV reverse transcriptase enzyme. Viral DNA development is stopped because the integrated nucleoside analogue lacks a 3′-OH group, which inhibits the formation of the 5′ to 3′ phosphodiester linkage necessary for DNA chain elongation 3-4.

Zidovudine, often known as azidothymidine or AZT, is an analog of thymidine. 1- [(2R,4S,5S)-4-azido-5-(hydroxyl methyl) oxolan-2yl] is its chemical name.Tetrahydropyrimidine-2,4-dione -5methyl-1,2,3,4-. It works by specifically blocking HIV’s overturn transcriptase, an enzyme the virus utilizes to replicate its RNA into DNA. The synthesis of HIV’s double-strands DNA, which is then incorporated into the inherited substance of the tainted cell, requires reverse transcription. AZT has a roughly 100-fold higher affinity for HIV’s overturn transcriptase, but at extremely high dosages, its triphosphate form may also block DNA polymerase, which is necessary for human cells to divide. It has been shown that the selectivity results from the cell’s capacity to swiftly fix its have DNA sequence in the event that AZT breaks it during synthesis, something that the HIV virus is unable to do. As a result, AZT prevents HIV replication while leaving uninfected cells functioning normally 5-6.

This review’s objectives are to provide an overview of the field’s history, explain the techniques used to find agents, and highlight potential antiviral medications that are already on the market and in development. Antivirals are especially used to treat viruses and highlight new and modern approaches to medication production as well as to stop the spread of viruses that are resistant to treatments.

Experimental Work

Chemicals and Reagents

Intas Pharmaceuticals Pvt. Ltd. provided the commercially available formulations of stavudine and zidovudine (API) as well as other necessary compounds for the study. We bought additional reagents from Rankem Ltd., including methanol, acetonitrile, HPLC water, buffer, orthophosphoric acid, triethylamine, 1N HCL, 1N NaOH, and 3% H2O2. Additionally, every reagent was of analytical quality 7.

Optimized Chromatographic state

HPLC testing circumstances were improved using the CosmosilC18 analytical column, which is 250 mm long, 4.6 mm wide, and has a particle size of 5μm.

Preparation of Solutions

Water containing orthophosphoric acid (0.1% V/V): Pipette 0.6 mL of orthophosphoric acid into a 500 mL measuring cylinder filled with 250 mL of HPLC-grade water, then top it over with the remaining 500 mL of HPLC mark water. Moved into a container of reagents and properly assorted the contents and kept at room heat. Three days after the preparation date, this solution was put to use 8.

Mobile phase: Acetonitrile: 0.1% OPA (54:44 % V/V)

450 mL of 0.1% OPA and 550 mL of aceto nitrile were taken out of a measuring cylinder, put into a reagent container, and properly assorted. Stored at room temperature after the preparation date, this solution was utilized within three days. This served as the diluent.

Stavudine Stock Solution, 4000 µg/ mL

40 mg of standard stavudine, precisely weighed, was put into a 10 mL vol. flagon, and the right quantity of methanol was further to get the ultimate stavudine concentration of 4000 µg/mL. The solution was utilized within seven days after the manufacture date and kept in a refrigerator at 5±3°C 9.

Zidovudine Stock Solution, 800 µg/ mL

8 mg of accurately measured standard Zidovudine was placed into a 10 mL vol. flask, and the suitable vol. of methanol was further to achieve the target closing attention of 900 µg/mL. The solution was utilized within seven days after the manufacture date and kept in a refrigerator at 5±3°C 10.

Combine Stock solution, (Stavudine 40 µg/mL and Zidovudine 9 µg/mL)

0.1 milliliters of hoard in a 10.0 mL volumetric flask, a solution of Stavudine 4000 µg/mL and Zidovudine 800 µg/mL was transferred. To get a solution with concentrations of Stavudine 40 µg/mL and Zidovudine 8 µg/mL, the volume was adjusted to the proper level using diluents. The solution was utilized within seven days after the preparation date and kept in a refrigerator at 5±3°C.

Method Validation

Linearity

By diluting the standard stock solution to provide a final concentration in the variety of 5–25 g/ml for Lamivudine and 10–50 gram/ml for Zidovudine for 5 gram/ml injection, linearity was achieved, and a calibration curve was created.

Precision (Repeatability)

In order to study the method’s intraday accuracy, the drug combination was repeatedly injected on the equal date. Following five measurements, The coefficient of variation was determined to be 5–25 µg/ml for stavudine and 10–50 µg/ml for zidovudine 11.

Precision (Intermediate):

Three duplicates of the standard concentration were injected by several analyzers to achieve intermediate accuracy. The percentage RSD was computed.

Accuracy

Three distinct concentrations of sample solutions were made: 80%, 100%, and 120%. A known quantity of sample was added to each solution, and the recovery of the additional sample was examined 12.

Robustness

The robustness of the RPHPLC method for the concurrent assessment of Stavudine and Zidovudine, along with their %RSD, was assessed by examining the effects of minor alterations in chromatographic conditions, including a pour rate variation of ±0.1 mL/min, a detection wavelength variation of ±1 nm, and a mobile phase variation of ±1 mL13.

Limit of quantification (LOQ)

LOQ is the lowly analyte deliberation that can be identified under the specified experimental circumstances with a reasonable level of precision and accuracy 14.

Result And Discussion

Assortment of immobile phase

Together Stavudine and Zidovudine are polar compounds, so analysed by means of a C18column.

Assortment of wavelength

Since the wavelength chosen affects the HPLC method’s selectivity, 265 nm was chosen as the detection wavelength for Stavudine and Zidovudine in order to ensure that both medications responded well.

Figure 1: Wavelength selection was based on the overlaid UV spectra of STV (20 µg/mL) and ZDV (4.5 µg/mL)

Click here to View Figure

Establishment of the retention time of individual drugs

To check for separation, a thinned assorted standard of the medicines was administered into the arrangement. The following parameters were ultimately determined to provide an optimum chromatogram after slight adjustments were made to the flow rate in order to increase resolution and the number of theoretical plates.

HPLC manufacturer: Agilent Laboratories 1100 slope arrangement

Particle size of stuffing: 5μm

The immobile stages: C18COSMOSIL (250mm x 4.6mm, 5 microns)

Detection wavelength: 283 nm

Flow rate: 0.7 mL/min

Movable Phase: Acetonitrile and 0.1% orthophosphoric acid (OPA) in a 54:44 %v/v ratio Temperature: Ambient

Injection Volume: 20 µL

The retention times and theoretical plate counts for both drugs are presented in Table 1, and the corresponding chromatogram is shown in Figure 2.

Table 1: Retention time of STV and ZDV

Parameter Stavudine Zidovudine Elapsed Retention Time (min) 3.668 5.925 Theoretical plates 5534 17824 Tailing factor 0.924 0.80 Declaration  – 10.8

System Suitability

For the conventional solutions, the tailing factor, resolution, and column efficiency were computed. The findings indicated that the approach was suitable for analysing the selected drug combinations. When the method is consistently executed, the system’s suitable parameters may reside inside a 2% relative variance range. Figure 3 shows the chromatogram, and Table 2 shows the system suitability parameter comparison findings.

Table 2: Proportional consequences of together the drugs

Parameter Stavudine Zidovudine Elapsed Retention Time (min) 3.75 5.99 Separation stage 5523.33 17823.17 Peak tailing ratio 0.91 0.80 Resolution 10.8 Figure 3: All scheme appropriateness variables of the developed process comply through its customary

Click here to View Figure

Method corroboration

Specificity

By injecting solutions of both medications, the method’s selectivity was examined. Sofosbuvir and Ledipasvir were found to have two distinct peaks at retention durations of 3.776 and 3.763, 6.004 and 5.993 minutes for the customary and illustration, correspondingly. The approach was specific since the drug standards plus the pharmaceuticals from model solutions had the identical retention periods (Figure 4).

Linearity

Five concentration levels were used to assess the method’s linearity. The response factor was graphed vs the drug concentration to establish the calibration curve. Stavudine and Zidovudine exhibit linearity within the attentiveness ranges of 39–199 μg/mL and 8–44 μg/mL, correspondingly. To get the calibration curve, construct a graph of Area against Concentration. Table 3 and the calibration curve in Figure 5 illustrate the linearity of Stavudine, whereas Table 4 and the calibration curvature in Figure 6 demonstrate the linearity of Zidovudine. The calibration curvature was establish to be linear.

Table 3: Linearity statistics of Stavudine

Conc. (μg/mL) Peak Area 1 Peak Area 2 Regular Peak area (n=2) 0 0 0 0 40 238.994 237.907 238.45 80 468.212 471.527 469.87 120 715.665 713.112 714.389 160 942.344 937.266 939.805 200 1200.008 1205.24 1202.624

Table 4: Linearity statistics of Zidovudine

Conc (μg/mL) Peak Area 1 Peak Area 2 Normal Peak area (n=2) 0 0 0 0 9 216.571 215.947 216.259 18 444.776 446.687 445.731 27 665.105 667.325 666.215 36 899.079 897.853 898.466 45 1139.157 1129.372 1134.264

Precision

The method’s accuracy was shown by research on structure repeatability (intraday precision) and transitional precision (inter-day precision) variations.

Scheme Repeatability (Intra – Day precision)

Six duplicates of the produced sample solutions were used to assess the repeatability of the system. For intra-day precision, the assay was performed six period at concentration level of 40,80, and 120 μg/ mL for Stavudine and 9, 8, and 27 μg /mL for Zidovudine. This allowed for the determination of climax area for the medications and the repeatability of sample application. RSD was computed and found to be less than 2% when peak locations were identified.

Transitional Precision (InterDay precision)

Three duplicates of the produced sample solutions were used to calculate the intermediate precision. The evaluation of three example sets conducted on distinct days at concentration level of 80 μg /mL for Sofosbuvir and 18 μg/ mL for Ledipasvir over various instance intervals to assess interday accuracy yielded the intermediate accuracy of sample application and peak area measurement. RSD was computed and found to be less than 2% when peak locations were identified.

Accuracy

By using the conventional addition procedure, which involves analyzing formulation samples to which specific quantities of genuine pharmaceuticals were added, the techniques’ correctness was guaranteed. The final mixes were tested, and the outcomes for both medications were contrasted with the predictions. Three separate quantities of standards (Stavudine 31,39,47 μg/ mL and Zidovudine 7.1,9,10.7μg/ mL) were spiked into beforehand analyze sample of the tablets (Stavudine 39 μg/ mL and Zidovudine 8 μg/ mL) in order to conduct the recovery tests in triplicate. The recovery investigations were conducted in compliance with ICH Guidelines, and the excellent recoveries using the standard addition technique demonstrate the good accuracy.

Robustness

Deliberate changes were made to the procedure parameters, including flow rate, pH, and wavelength, in order to assess the robustness of the suggested approach. It was found that the method’s analytical performance did not significantly alter. According to the findings, the technique’s robustness was shown by the short values of % RSD (< 2) of % drug substance that were achieved after making minor adjustments to the method parameters.

Limit of Detection and Limit of Quantification

The ICH Guidelines were adhered to in establishing the limits of detection (L O D) and quantitation (L O Q) for both stavudine as well as zidovudine. The limit of detection (L O D) was considered as 3.3σ/S and the limit of quantification (L O Q) as 10σ/S, base on the typical divergence of the answer and grade of the calibration curvature established in a short concentration range of the objective analyte. S represents the slope of the calibration curve, whereas σ (the standard divergence of the rejoinder) was unwavering using the standard divergence of the y-intercepts of the deterioration lines. 

Table 5: Limit of Detection (LOD) and Limit of Quantification (LOQ)

S. No. DRUG LOD (µg/mL) LOQ (µg/mL) 1 Stavudine 0.420 1.27 2 Zidovudine 0.051 0.177

 Assay: Established HPLC technique was used for the examination of a commercial tablet formulation.

Table 6: % Results of Assay of marketed formulation

Marketed Formulation Ingredients Conc. µg/mL Area (n=3) Amount Found (µg/mL) % Assay Harvoni Stavudine 80 460.86 78.36 98.20% Zidovudine 18 452.285 17.284 100.58%

Conclusion

Goal of the current effort is to develop the RP-HPLC technique and optimize the chromatographic conditions. The suggested approach may be effectively used for the assessment of stavudine and zidovudine in pharmaceutical dose forms since it is quick, easy, sensitive, accurate, robust, and exact. The percentage RSD from five duplicate injections for Zidovudine and Stavudine retention durations and peak areas was used to assess the system appropriateness characteristics using standard chromatograms. Lamivudine and zidovudine were found to have linearity and correlation coefficients of 0.999. It was done with precision.

References

Chatwal G.R, Sharma A, Instrumental method of chemical Analysis, Himalaya Publishing House Delhi, 5th Edn.2004.1.1-1. Kasure A.V., Mahadik K.R., More H.N., Instrumental Methods 14(2), Nirali Prakashan, Pune 2006 ,44-55. Skoog DA, Holler FJ, Nieman TA, Principles of Instrumental Analysis Harcourt Asia Pvt.Ltd. & Harcourt Collage Publishers. 1998.329-335. Christan GD., A text book of analytical chemistry; john wiley and sons, inc, 2004, 2- 4. Skoog., Leauary J., A text book of principle of instrumental analysis;54th Edn . Singapor: Thomsons asian Pvt Ltd; 2004. 3 Skoog D., Holler F., Timothy A., Nieman N., A Text book of principles of instrumental analysis; 4th Edn. Londan: sanders college publications; 1999. 1-2, 338-340. V arma RM., A text book of Analytical chemistry; 3rd Edn. new Delhi: CBS publishers and distributors; 2007. 3-10. Chatwal GR., Anand AK., A text book of instrumental methods of chemical analysis; Himalaya publishing house, 2007, 1.3-1.7. Chatwal GR., Anand SK., A text book of instrumental methods of chemical analysis; Himalaya publishing house; 2007. 2.566-2.568 Skoog DA., West DM., Holler FT.,A text book of Fundamentals of Analytical Chemistry; 7th Edn. Singapore: Thomsan Asia Pvt. Ltd.; 2001. 725-726. Sethi PD., A text book of HPLC Quantitative Analysis of Pharmaceutical Formulations; 1st Edn. New Delhi: CBS Publishers and Distributors; 2001; 1-28. Mendham J., Denny RC., Barnus JD., Thomsons N., A text book of Vogel quantitative analysis of drug in pharmaceutical analysis; pearson Edn. Singapore: 2003. 8-9. Sethi PD., A text book of HPLC Quantitative Analysis of Pharmaceutical Formulations; 1st Edn. New Delhi: CBS Publishers and Distributors; 2001. 52-53.11-19 Sharma BK., A text book of instrumental method of chemical analysis; merat goal publishing house: 2006. c-288-c-302